![ZOLL X Series Operator'S Manual Download Page 107](http://html1.mh-extra.com/html/zoll/x-series/x-series_operators-manual_4055648107.webp)
Configuring Heart Rate (HR) Meter Alarms
9650-001355-01 Rev. M
X Series Operator’s Guide
7-7
Life Threatening Rhythm Alarms
When LTA monitoring is enabled, the X Series unit will monitor for the following life
threatening ECG rhythms: asystole, ventricular fibrillation, ventricular tachycardia, extreme
bradycardia, and extreme tachycardia. These events are also displayed on Treatment Summary
Reports, Alarm Snapshots, and Continuous Waveform printouts.
The following example shows an Asystole alarm event:
To enable or disable LTA alarms and set Extreme Brady/Tachy alarm limits, you can either do
so through the Alarms
quick access key (
), or the
HR/PR Parameter Control Panel
.
To configure LTA alarms through the Alarms quick access key:
1. Press .
2.
Press .
3.
Press
. Use the navigation keys to highlight and select the fields that you want to change:
•
LTA Monitoring
•
Extreme Brady/Tachy Alarms
When you are finished changing values on the alarm menu, navigate to the Backarrow key to
confirm your choices and exit the menu.
Limits
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...