Battery
•
Although the device can operate with auxiliary power alone, ZOLL strongly recommends
that you operate the unit with a battery installed at all times. Operating the unit with a
battery provides a backup in case of AC power shortage, and results in faster charge time.
The battery can be automatically recharged while it is installed in the unit. Keep a fully
charged spare battery pack with the defibrillator at all times.
•
Test battery packs regularly. A battery that does not pass the ZOLL charger’s capacity test
might cause the X Series unit to shut down unexpectedly.
•
If the Low Battery indication occurs at any time during operation, immediately replace the
battery pack.
•
If the
LOW BATTERY
icon appears, plug the X Series unit into a power source or install a
fully charged battery pack. When the warning low battery shutdown prompt appears,
immediately replace the battery pack with a fully charged pack or plug the X Series unit into
a power source, as unit shut down due to a low battery condition is imminent.
•
If mistreated, a battery pack might explode. Do not disassemble a battery pack or dispose of
it in fire.
Operator Safety
•
The X Series can deliver 200 joules of electrical energy. If this electrical energy is not
discharged properly, as described in this manual, the electrical energy could cause personal
injury or death to the operator or bystanders.
•
Do not use the X Series in the presence of oxygen-rich atmospheres, flammable anesthetics,
or other flammable agents (such as gasoline). Using the unit in such environments might
cause an explosion.
•
Do not use the unit near or within standing water. Electrical safety might be compromised
when the defibrillator is wet.
•
Never discharge the unit with the defibrillation electrodes or paddles shorted together or in
open air.
•
Do not discharge the defibrillator except as indicated in the instructions. Discharge the
defibrillator only when defibrillation electrodes or paddles are properly applied to the
patient.
•
To avoid risk of electric shock, this equipment must only be connected to AC mains with
protective earth
•
To avoid risk of electrical shock, do not touch the gelled area of the hands-free therapy
electrodes during pacing or defibrillation.
•
To avoid risk of electrical shock, do not allow electrolyte gel to accumulate on hands or
paddle handles.
•
To avoid risk of electrical shock, do not allow patient connectors to contact other conductive
parts, including earth.
•
For defibrillation using paddles, use only high-conductivity electrolyte gel specified for
such use by the manufacturer.
•
When using paddles for defibrillation, use your thumbs to operate the
SHOCK
buttons.
Doing so avoids inadvertent shock to the operator.
•
The use of accessory equipment that does not comply with the equivalent safety
requirements of the X Series defibrillator could reduce the level of safety of the combined
system. When choosing accessory equipment, consider the following:
•
Use of the accessory in the patient vicinity.
•
Evidence that the safety certification of the accessory has been performed in accordance
with the appropriate IEC (EN) 60601-1 and/or IEC (EN) 60601-1-1 harmonized
national standards.
•
Always check that the equipment functions properly and is in proper condition before use.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...