Cautions
9650-001355-01 Rev. M
X Series Operator’s Guide
10-5
•
To protect from electric shock, always remove the sensor and completely disconnect the
pulse co-oximeter before bathing the patient.
•
SpCO readings may not be provided if there are low arterial saturation levels or elevated
methemoglobin levels.
•
SpO2, SpCO, SpMet, and SpHb are empirically calibrated in healthy adult volunteers with
normal levels of carboxyhemoglobin (COHb) and methemoglobin (MetHb).
•
Do not adjust, repair, open, disassemble, or modify the pulse co-oximeter or accessories.
Injury to personnel or equipment damage could occur. Return the pulse co-oximeter for
servicing if necessary.
Cautions
•
Do not place the pulse co-oximeter where the controls can be changed by the patient.
•
Electrical shock and flammability hazard: Before cleaning, always turn off the instrument
and disconnect from any power source.
•
When patients are undergoing photodynamic therapy they may be sensitive to light sources.
Pulse oximetry may be used only under careful clinical supervision for short time periods to
minimize interference with photodynamic therapy.
•
Do not place the pulse co-oximeter on electrical equipment that may affect the instrument,
preventing it from working properly.
•
If SpO2 values indicate hypoxemia, a laboratory blood sample should be taken to confirm
the patient’s condition.
•
If the Low Perfusion message is frequently displayed, find a better perfused monitoring site.
In the interim, assess the patient and, if indicated, verify oxygenation status through other
means.
•
If using pulse oximetry during full body irradiation, keep the sensor out of the radiation
field. If the sensor is exposed to the radiation, the reading might be inaccurate or the
instrument might read zero for the duration of the active irradiation period.
•
The instrument must be configured to match your local power line frequency to allow for
the cancellation of noise introduced by fluorescent lights and other sources.
•
To ensure that alarm limits are appropriate for the patient being monitored, check the limits
each time the pulse co-oximeter is used.
•
Variation in hemoglobin measurements may be profound and may be affected by sampling
technique as well as the patient's physiological conditions. Any results exhibiting
inconsistency with the patient’s clinical status should be repeated and/or supplemented with
additional test data. Blood samples should be analyzed by laboratory instruments prior to
clinical decision making to completely understand the patient’s condition.
•
Do not submerge the pulse co-oximeter in any cleaning solution or attempt to sterilize by
autoclave, irradiation, steam, gas, ethylene oxide or any other method. This will seriously
damage the pulse co-oximeter.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...