Chapter 9 Monitoring CO2
9-2
www.zoll.com
9650-001355-01 Rev. M
Warning!
•
When using a sampling line for intubated patients with a closed suction system, do
not place the airway adapter between the suction catheter and endotracheal tube.
This is to ensure that the airway adapter does not interfere with the functioning of
the suction catheter.
•
Do not cut or remove any part of the sample line. Cutting the sample line could lead
to erroneous readings
•
If too much moisture enters the sampling line (i.e., from ambient humidity or
breathing of unusually humid air), the message Clearing FilterLine will appear in
the message area. If the sampling line cannot be cleared, the message
FilterLine
Blockage
will appear in the message area. Replace the sampling line once the
FilterLine Blockage
message appears.
•
Do NOT use the Microstream sidestream CO
2
accessories in the presence of
flammable anesthetics or other flammable gases.
•
Loose or damaged connections may compromise ventilation or cause an inaccurate
measurement of respiratory gases. Securely connect all components and check
connections for leaks according to standard clinical procedures.
•
Always ensure the integrity of the patient breathing circuit after insertion of the
airway adapter by verifying a proper CO
2
waveform (capnogram) on the monitor
display.
Caution
Microstream
®
EtCO
2
sampling lines are designed for single patient use, and are not to be
reprocessed. Do not attempt to clean, disinfect, sterilize or flush any part of the sampling line as
this can cause damage to the monitor.
Dispose of sampling lines according to standard operating procedures or local regulations for the
disposal of contaminated medical waste.
Before use, carefully read the Microstream EtCO
2
sampling line’s
Directions for Use
.
CO
2
Monitoring Setup and Use
Follow these steps to set up CO
2
monitoring:
1. Select the correct CO
2
sampling line for the patient.
2.
Attach the sampling line to the unit’s CO
2
inlet port.
3.
Apply the Filterline airway adaptor or Smart CapnoLine Nasal or Nasal/Oral cannula to the
patient.
4.
Check that the X Series unit is set up for the correct patient type -- Adult, Pediatric, or
Neonate.
5.
Configure alarms (if the current alarm settings are not appropriate) and other CO
2
features.
6.
Press the CO
2
quick access key (
)to initiate CO
2
monitoring.
CO
2
CO
2
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...