Rezeroing a Transducer
9650-001355-01 Rev. M
X Series Operator’s Guide
11-5
Rezeroing a Transducer
You can re-zero a transducer at any time by opening the transducer stopcock to atmospheric air.
If the unit accepts the new zero reference value, the unit displays pressure values that are based
on that value and adjusts the waveform according to the new scale.
Warning!
If you attempt to zero an IBP channel after the channel has been successfully zeroed
and is currently monitoring a pressure waveform, the unit will display the message,
ZERO REJECTED
, in the IBP channel’s numeric window. This message will preempt
the display of the valid invasive pressure numerics.
Displaying IBP Measurements
After attaching and zeroing a transducer, the X Series unit displays the invasive pressure’s
systolic, diastolic, and MEAN values in the IBP channel’s numeric display and, optionally (if
enabled through the unit’s Waveform Select menu) the waveform for that IBP channel:
The X Series unit allows you to specify a label that identifies the channel’s IBP measurement,
and to select a display format for the numeric display.
The unit displays the waveform scales after you zero the transducer. When the unit accepts the
zero reference value, it determines the waveform scales and displays them.
Conditions Affecting IBP Measurements
When reading blood pressure measurements on the IBP numeric display, keep in mind that the
following conditions can affect the accuracy of IBP measurements:
•
Catheter placement in the vasculature. Artifact such as catheter whip should be handled per
your established clinical protocols.
•
Position of the transducer stopcock, catheter, and flush port.
•
Saline line flushes which will temporarily interrupt accurate pressure measurement.
•
Position of the transducer relative to the patient’s phlebostatic axis or catheter tip.
•
Patient movement.
•
Catheter clogging.
•
Air bubbles in catheter or transducer dome.
Caution
Flush catheter regularly while taking IBP measurements. Always view the IBP waveform to
ensure that pressure measurements are based on a physiological waveform.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...