Supervisor Communications Menu
9650-001355-01 Rev. M
X Series Operator’s Guide
22-31
To configure clock synchronization:
1. Under the Communications menu, select
Clock Sync > Configure.
The Clock
Synchronization menu displays.
2.
In the
Clock Sync
field, specify whether or not to synchronize the X Series internal real time
clock to an external time source.
—
Select
SNTP
if you would like to synchronize the X Series clock to an external time
source.
—
Select
None
(default) if you do not wish to synchronize the X Series internal real time
clock to an external source.
3.
In the
Time Server Name
field, specify the time server to use for clock synchronization by
either entering the host name or an IPv4 address for the server.
4.
In the
Time Server Port
field, enter a numeric UDP port number for connecting to the
external server. The default is 123.
5.
In the
Daylight Saving Time
field, indicate whether to use Daylight Saving Time or not by
selecting
On
or
Off
(default) as appropriate. Adjustments to Daylight Saving Time take
effect immediately and must be manually adjusted at every transition between Standard Time
and Daylight Saving Time. The Daylight Saving Time offset is zero if Daylight Saving Time
is set to
Off
and 60 minutes if Daylight Saving Time is set to
On
.
Note:
UTC time is a constant reference time. It does not change to reflect Daylight
Saving Time or Summer Time. The Daylight Saving Time field allows you to
accommodate locales that observe Daylight Saving Time or Summer Time. So, for
example, UTC is 5 hours ahead of local time on the east coast of the United States
during winter, but 4 hours ahead during summer.
6.
Indicate the time zone offset value to use for your particular locale in the
Time Zone Offset
Hours
(default 0) and
Time Zone Offset Mins.
(default 0) fields as appropriate.
When you are done entering data, press the back arrow (
) to return to the Supervisor >
Communications menu. When changes are made on the Clock Sync menu the X Series attempts
to synchronize its time if a network connection is available. To ensure the real time clock never
changes during patient treatments, any necessary clock synchronization corrections will not
occur, however, until the unit has been off for at least two minutes and prior to the start of a new
case.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...