Battery Status and Auxiliary Power Indicators
The battery status indicator displays various battery icons to indicate the approximate
remaining unit run time based on the charged state of the battery. Additionally, these icons
provide indications of the status of the battery connection and communication with the unit.
The auxiliary power indicator indicates that the unit is being powered by the auxiliary power
adapter.
Note:
Upon powering up the X Series unit, the battery capacity will be displayed within
approximately 15 seconds under normal conditions. Under some circumstances, such
as activating the defibrillator immediately after the unit is turned on, the battery icon
may display less than one hour battery capacity for up to two minutes after exiting the
defibrillation mode.
Icon
Status
Indication/Action
Auxiliary power adapter is
connected
The unit is being powered by the
auxiliary power adapter.
No battery detected
Either there is no battery in the unit
while it is being powered by the
auxiliary power adapter, or the
device cannot detect that the
battery is connected.
Low battery capacity
Replace the battery soon.
Communication failure
The unit is unable to establish
communication with the battery
and the battery capacity is
unknown. Check the battery
contacts.
Battery fault
A battery fault has been detected.
Replace the battery.
Battery Level 1
The battery has less than one hour
of remaining battery capacity.
Battery Level 2
The battery has greater than one
hour of remaining battery capacity.
Battery Level 3
The battery has greater than two
hours of remaining battery
capacity.
Battery Level 4
The battery has greater than three
hours of remaining battery
capacity.
Battery Level 5
The battery is fully charged.
Low
1:00+
2:00+
3:00+
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...