Pacer
9650-001355-01 Rev. M
X Series Operator’s Guide
A-19
DC Auxiliary Operating Power
:
8300-000006
Input: 12-24V
11A
Output: 14.45V
4.15A
80W peak
IP Rating: IP24
Pacer
Type
: External transcutaneous pacing
Pacer Rate
: 30 to 180 BPM ± 1.5%.
Output Current
: 0 to 140 mA ± 5% or 5 mA (whichever is greater)
Modes
: Demand and Fixed
Status Indicators
:
ECG lead fault, pace marker on monitor and chart, start/stop indicator on
display.
Pulse Type
: Rectilinear, constant current
Pulse Width
: 40 ms +/-2 ms
CO
2
Range
: 0 to 150 mmHg
Accuracy CO
2
:
* At sea level.
** Accuracy applies for breath rates of up to 80 bpm. For breath rates above 80 bpm,
accuracy is 4 mmHg or ±12% of reading whichever is greater, for EtCO
2
values exceeding
18 mmHg. This is tested according to and is compliant with ISO 21647. To achieve the
specified accuracies for breath rates above 60 breaths/minute, the Microstream FilterLine
H Set for Infant/Neonatal must be used. Above 40 C, ± 1mmHg or ± 2.5% (whichever is
greater) has to be added to the tolerance of the accuracy specs.
CO2 Sampling Rate:
50 msec
Drift of Measurement Accuracy
: Over any 24-hour period, the accuracy claims listed above
are maintained.
The accuracy specification is maintained to within 4% for the following gas mixtures (all
values are in Vol.%).
CO
2
Partial Pressure*
Accuracy **
0-38 mmHg
± 2 mmHg
39-99 mmHg
± (5% of r 0.08% for every 1 mmHg above 38 mmHg)
100-150 mmHg
± (5% of r 0.08% for every 1 mmHg above 38 mmHg)
CO
2
N
2
O
2
N
2
O
H
2
0
Anesthetic Agents
1.0 to 13 0 to 97.5
0 to 100
0 to 80
Dry to saturated
According to EN 21647
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...