Using the SpO2 Parameter Control Panel
9650-001355-01 Rev. M
X Series Operator’s Guide
10-15
Selecting the SpO
2
Sensitivity
You can select either normal or high sensitivity for SpO
2
monitoring. The Normal sensitivity
setting is recommended setting for most patients. The High sensitivity setting allows you to
monitor SpO
2
even under very low perfusion conditions. Such conditions may include severe
hypotension or shock. With the High sensitivity setting, however, SpO
2
results are more easily
contaminated by artifact. To ensure accurate SpO
2
readings when using the High sensitivity
setting, observe the patient carefully and continuously.
Selecting SpHb Venous Mode
If the SpHb option is installed, the X Series unit allows you to specify whether to use venous
mode (
On
) as the blood sample source or not (
Off
). When
Off
(default) the X Series uses
arterial as the blood sample source.
Specifying the SpHb Averaging Time
If the SpHb option is installed, the X Series unit allows you to choose the time period over
which SpHb values will be calculated:
Short
,
Medium
, and
Long
(default). Averaging time
represents the approximate time period the X Series will use to calculate SpHb values. Longer
averaging times allow you to see subtle changes in the SpHb value and tend to improve
accuracy.
Selecting the Heart Rate/ Pulse Rate (HR/PR) Tone
The unit allows you enable or disable the tone that the monitor uses to indicate detection of the
patient’s pulse:
On
or
Off
(no tone sounds). The default tone is
Off
.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...