X Series Product Functions
9650-001355-01 Rev. M
X Series Operator’s Guide
1-13
Batteries
X Series models use an easily replaced rechargeable lithium-ion battery pack (the
SurePower II
Battery Pack). A new, fully charged battery pack typically delivers more than 6 hours of ECG
monitoring. Use of other functions (such as the defibrillator, printer, or pacemaker) reduces this
time.
When a
LOW BATTERY
icon appears on the display and the unit emits three beeps in
conjunction with the displayed battery icon, the battery must be replaced and recharged.
You can charge the battery by either of the following methods:
•
Internal charging
— plug the X Series into an auxiliary power adapter to automatically
begin charging the installed battery pack. The front panel battery indicator operates as
follows:
Note:
Upon power up, it takes approximately 45 seconds for the LEDs on the battery to
accurately display run time.
•
External charging
— use the ZOLL SurePower Battery Charger with the X Series battery
adapter to charge the battery pack and test the battery’s capacity. For details, refer to the
SurePower II Battery Pack Guide
.
The Recalibration LED icon ( ) lights for approximately 10 seconds (after you press and
release the Display button) if the battery needs to be calibrated. If the Recalibration LED lights,
the runtime indicator will not display run time for that battery. For best performance of the
battery, you should recalibrate the battery as soon as possible.
To manually recalibrate the SurePower II Battery Pack, you can insert the battery into the
SurePower Charger Station and perform a Manual Test (for more information, see the ZOLL
SurePower
Charger Station Operator’s Guide
).
After you recalibrate the battery, the Recalibration LED will only flash when you press the
Display button.
When the indicator is:
It means:
Steady yellow
Battery is charging.
Steady green
Battery is charged.
Alternating yellow and
green
The charge state cannot be
determined or a battery charging
fault has been detected.
Not lit
No battery in device.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...