Electromagnetic Compatibility Guidance and Manufacturer’s Declaration
9650-001355-01 Rev. M
X Series Operator’s Guide
A-33
Electromagnetic Immunity: Non Life-Supporting Functions
Non life-supporting functions of the X Series include
:
12-lead ECG, EtCO
2
, SpO
2,
SPCO,
SpMet, NIBP, IBP, Temp, Impedance Respiration
.
Guidance and manufacturer’s declaration – electromagnetic immunity – for non–
life supporting equipment and systems
The non life-supporting functions of the X Series are intended for use in the electromagnetic environment
specified below. The customer or user of the X Series should ensure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment – guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the X Series, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
a
3 Vrms
d
= 1.2
Radiated RF
IEC 61000-4-3
(IBP)
Radiated RF
IEC 61000-4-3
(all other non-life
supporting
functions)
3 V/m
80 MHz to 2.5 GHz
3 V/m
80 MHZ to 2.5 GHz
3 V/m
10V/m
d
= 1.2
80 MHz to 800 MHz
d
= 2.3
800 MHz to 2.5 GHz
d
= 0.35
80 MHz to 800 MHz
d
= 0.7
800 MHz to 2.5 GHz
where
P
is the maximum output power rating
of the transmitter in watts according to the
transmitter manufacturer and
d
is the
recommended separation distance in meters.
Field strengths from fixed RF transmitters, as
determined by electromagnetic site survey,
a
should be less than the compliance level in
each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the X Series is used exceeds the applicable RF
compliance level above, the X Series should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the X Series.
b. Over the frequency ranges 150 kHz to 80 MHz field strength should be less than 3 V/m.
P
P
P
P
P
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...