Impedance Pneumography
9650-001355-01 Rev. M
X Series Operator’s Guide
A-15
Heart Rate Response to Irregular Rhythm
: (AAMI EC13-2002, section 4.1.2.1.e.)
Ventricular Bigeminy
: 80 BPM (expected)
Slow Alternating Ventricular Bigeminy
: 60 BPM (expected)
Rapid Alternating Ventricular Bigeminy
: 120 BPM (expected)
Bidirectional Systole
: 45 BPM (expected)
Tachycardia Response Time
:
Response time to tachycardia alarm is on average 3.4 seconds per AAMI EC-13-2002,
section 4.1.2.1.g, and IEC 60601-2-27:2011, subclause 201.7.9.2.9.101 b) 6). Response
times include a 1.0 second display update interval.
Pacemaker Pulse Rejection:
(In accordance with AAMI EC13:2002, section 4.1.4 and IEC 60601-2-27:2011, subclause
201.12.1.101.13)
•
Pulses without overshoot: Rejects all pulses with amplitude of +2 mV to +700 mV and
duration of 0.1 ms to 2 ms, with no tail.
•
Pulses with overshoot: Rejects all pulses with amplitude of +2 mV to +700 mV and duration
of 0.1 ms to 2 ms, with overshoot up to 100 ms.
•
A-V sequential pulses: A-V sequential pacemaker pulses may not be rejected.
•
Fast ECG signals: Approximately 50% of ECG input pulses with a slew rate of 3 V/s RTI
may trigger the pacemaker pulse detector.
Electrosurgery Protection
: The X Series is suitable for use in the presence of electrosurgery
as specified in IEC 60601-2-27. Burn hazard protection via a 1K current limiting resistor
contained in each ECG leadwire.
Impedance Pneumography
Displayed Data
: Numeric breath rate, Impedance waveform
Breath rate range
:
Adult, Ped: 2 to 150 breaths / minute
Neonates: 3 to 150 breaths / minute
Breath rate accuracy
: 2% or +/- 2 breaths / minute, whichever is greater
Displayed Breath Rate:
Average of last 10 breath-to-breath rates.
Leads
: Lead I (RA – LA), Lead II (RA – LL)
Sweep Speed
: 3.13, 6.25, 12.5 mm/sec
Alarm settings
: High, low and no breath rate alarm
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...