Synchronized Cardioversion
9650-001355-01 Rev. M
X Series Operator’s Guide
15-11
Synchronized Cardioversion
Warning!
Only skilled personnel trained in Advanced Cardiac Life Support and familiar with
equipment operation should perform synchronized cardioversion. The precise cardiac
arrhythmia must be determined before attempting defibrillation or cardioversion.
Before attempting synchronized cardioversion, ensure that ECG signal quality is
sufficient to minimize the risk of synchronizing on artifact.
Certain arrhythmias, such as ventricular tachycardia, atrial fibrillation, and atrial flutter, require
synchronizing the defibrillator discharge with the ECG R-wave to avoid the induction of
ventricular fibrillation. In this case, a synchronizing (SYNC) circuit within the defibrillator
detects the patient’s R-waves. When the
SHOCK
button (or buttons, if using paddles) is
pressed and held, the unit discharges with the next detected R-wave, thus avoiding the
vulnerable T-wave segment of the cardiac cycle.
When in the SYNC mode, the unit displays markers (
S
) above the ECG trace to indicate the
points in the cardiac cycle (R waves) where discharge can occur.
Verify that markers are clearly visible on the monitor and their location is appropriate and
consistent from beat to beat.
The synchronized cardioversion procedure for ZOLL hands-free therapy electrodes is identical
to that for paddles with the exception of the
SHOCK
button location.
S marker indicates each detected
R wave during synchronization.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...