Chapter 11 Monitoring Invasive Pressures (IBP)
11-6
www.zoll.com
9650-001355-01 Rev. M
Enabling/Disabling IBP Alarms and Setting Alarm Limits
When enabled, the X Series unit sounds alarms whenever IBP measurements are outside set
limits for the following:
•
High and Low Systolic Pressure
•
High and Low Diastolic Pressure
•
High and Low Mean Arterial Pressure (MEAN)
Note:
When enabled, any of the low alarms for IBP (systolic, diastolic, or mean) will also
provide an alert in the event of a disconnected catheter.
To enable (or disable) IBP alarms and set Upper and Lower alarm limits and IBP channel, you
can either use the navigation buttons to highlight and select the IBP channel display or:
1. Press the More
quick access key (
).
2.
Press .
3.
Press the
Limits
quick access key (
).
4.
Use the navigation buttons to highlight and select the appropriate alarm menu selection. For
each IBP channel, there are alarm menus for systolic (Systolic Alarm), diastolic (Diastolic
Alarm) and mean arterial pressure (Mean Alarm) alarm settings. On the selected Alarm
Settings menu, use the navigation keys to select the fields that you want to change. The fields
are
•
Upper Limit Enable
•
Lower Limit Enable
•
Upper Limit
•
Lower Limit
5.
When you are finished changing values on the alarm
menu, navigate to the Backarrow key
to exit the menu.
Setting Upper and Lower Systolic (SYS) Alarm Limits
Initially, the IBP Systolic Alarm Settings
menu specifies that the IBP systolic pressure alarms
are enabled (
ON
) or disabled (
OFF
)
,
and displays the default upper and lower systolic limits.
The upper and lower limits can be
ON
or
OFF
(default is
OFF
). The following table lists the
default IBP Systolic limits for adult, pediatric, and neonate patients, and gives the range in
which you can set these limits:
Limits
Patient Type
IBP Systolic Limit Default
IBP Systolic Limit Range
Adult
Lower: 75 mmHg
Upper: 220 mmHg
Lower: -30 to 298 mmHg
Upper: -28 to 300 mmHg
Pediatric
Lower: 75 mmHg
Upper: 145 mmHg
Lower: -30 to 298 mmHg
Upper: -28 to 300 mmHg
Neonate
Lower: 50 mmHg
Upper: 100 mmHg
Lower: -30 to 298 mmHg
Upper: -28 to 300 mmHg
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...