Defibrillator/Pacing Test with Hands-Free Therapy Electrodes
9650-001355-01 Rev. M
X Series Operator’s Guide
24-3
Defibrillator/Pacing Test with Hands-Free Therapy
Electrodes
Note:
If a low battery message appears during any of this testing, the battery is close to
depletion and should be replaced or recharged.
Warning!
Keep hands, fingers, and other conductive materials away from paddle electrode plates
when discharging the defibrillator, or pacing.
Function
Response
1
Connect the external AC adapter to a working
AC outlet and to the X Series rear panel.
Verify that the green Auxiliary Power LED
illuminates on the X Series front panel.
2
Insert a battery into the unit (if a battery is not
already in the slot).
Verify that the Battery Charge LED on the X
Series front panel illuminates yellow. When the
battery pack has been fully charged, the
battery status indicator light is green.
3
Press the unit's power on switch.
Verify the unit issues two audio beeps and that
the red, yellow, and green Visual Alarm
indicators illuminate for 2 to 3 seconds.
Note
: The yellow LED may remain lit after the
unit powers on.
The X Series unit performs a number of power
on self tests.
Verify that the unit displays
SELF TEST
PASSED
.
If the message
SELF TEST FAILED
appears
on the display, contact the appropriate
technical personnel or the ZOLL Technical
Service Department.
4
Disconnect the external AC adapter from the
X Series rear panel.
Verify that the unit continues to operate without
interruption on battery power and that the
displayed battery icon indicates that a fully
charged battery is installed.
Verify that no error messages are displayed.
5
With no ECG cables attached to the unit,
select Lead I, II, or III as the waveform display
source.
Verify that a
LEAD FAULT
message displays in
the waveform window.
6
Select
Pads
as the waveform display source
and connect the multifunction cable to the
unit. (Do not connect the test connector to the
MFC.)
Verify that a
CHECK THERAPY
ELECTRODES
message displays in the
waveform window.
7
Connect the multi-function cable (MFC) to the
test connector.
Verify that the
CHECK THERAPY
ELECTRODES
message is replaced by a
SHORT DETECTED
message.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...