ECG Monitoring Setup
9650-001355-01 Rev. M
X Series Operator’s Guide
6-3
ECG Monitoring Setup
The proper application and placement of electrodes is essential for high quality ECG
monitoring. Good contact between the electrode and skin minimizes motion artifact and signal
interference.
The following procedure describes how to monitor a patient’s ECG using 3- and 5-Lead ECG
cables. For information on the application and use on MultiFunction Pads and External
Paddles, which you can also use to monitor ECG, see Chapter 15,
Manual Defibrillation.
To monitor a patient’s ECG using 3- and 5-Lead ECG cables, perform the following steps:
1. Prepare the patient's skin for electrode application:
2.
Apply the electrode pads to the patient.
3.
Connect each lead of the ECG cable to the appropriate electrode.
4.
Insert the patient cable plug into the ECG input connector on the X Series unit.
5.
Select the ECG waveforms to be displayed on the waveform trace display screen.
6.
Observe the patient‘s electrocardiogram on the display, and adjust size of the ECG waveform
trace, as necessary.
Preparing the Patient for Electrode Application
The proper application of electrodes is essential for high quality ECG monitoring. Good contact
between the electrode and skin minimizes motion artifact and signal interference.
Before applying electrodes, prepare the patient’s skin, as necessary:
•
Shave or clip excess hair at electrode site.
•
Clean oily skin with an alcohol pad.
•
Rub site briskly to dry.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...