Chapter 11 Monitoring Invasive Pressures (IBP)
11-2
www.zoll.com
9650-001355-01 Rev. M
Warning!
•
If electrocautery is used, always avoid using any transducer with a conductive
(metal) case connected to its ground shield. Using a conductive transducer case that
is connected to its cable shield risks high-frequency burns at the ECG electrodes if
the transducer case becomes grounded to earth.
•
Normal alarm functions will detect complete disconnections of invasive pressure
transducers; however, the alarm functions will not detect a partial disconnection or
the use of some incompatible transducers. Use only approved transducers and ensure
that the transducers are connected properly.
•
Before you use the X Series unit on a new patient, always turn it off for at least 2
minutes. This clears the previous patient’s trend values, alarm limit settings, and
NIBP cuff inflation pressure.
•
Use
only
ZOLL-approved IBP sensors. Use of non-approved sensors may result in
inaccurate IBP measurements.
IBP Setup
To take safe and accurate IBP measurements using the X Series, you must perform the
following steps, each of which corresponds to a section in this chapter. Read each section
carefully before you perform IBP measurements.
1. Attach the invasive pressure transducer to the X Series unit.
2.
Zero the transducer.
3.
Set the invasive pressure alarms (according to your organization’s standards).
4.
Select a label for the invasive pressure channel.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...