AED Operation
9650-001355-01 Rev. M
X Series Operator’s Guide
13-5
2 Analyze
WARNING! Do not analyze the patient ECG during patient movement. A patient must be motionless
during ECG analysis. Do not touch the patient during analysis. Cease all movement via
stretcher or vehicle before analyzing the ECG.
The X Series automatically begins the analysis of the patient’s ECG rhythm,
displays an
ANALYZING ECG message for 5 seconds, and announces and displays a
STAND CLEAR
message. If therapy electrodes have not been properly connected to the patient, an
ATTACH
PADS
or
CHECK PADS
message is displayed and analysis will be inhibited.
Note:
You can press
Pause
to temporarily stop the rescue cycle. While the rescue cycle is
paused, ECG analysis continues in the background. The trends, log, alarm monitoring,
and 12-lead features are only accessible when the rescue cycle is paused. Press
Analyze
to resume the analysis process.
Note:
If the X Series has been configured to perform extra CPR at startup, it displays a
CHECK PULSE
voice prompt and the message appears for 10 seconds. Then it
displays a
PERFORM CPR
message along with a voice prompt for the configured
duration before analysis begins. You can start an ECG analysis during the CPR interval
by pressing the
ANALYZE
button.
An ANALYZING ECG message is displayed while the patient’s ECG is analyzed. Once the
analysis is completed, the unit indicates whether or not a shock is advised.
WARNING! ECG rhythm analysis does not warn of patient asystole, which is not a shockable
rhythm.
STAND
CLEAR
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...