![background image](http://html1.mh-extra.com/html/zoll/x-series/x-series_operators-manual_4055648294.webp)
Chapter 24 Maintenance
24-4
www.zoll.com
9650-001355-01 Rev. M
8
Press the Select Energy arrow (up or down).
Verify that the defibrillation window displays
and the following message appears:
PADS SHORT DETECTED
SELECT 30J TO TEST
9
Press the Select Energy arrow (up or down)
to select 30 Joules on the X Series unit.
Verify that the defibrillator window shows
30 J as the selected energy.
10 Press the
CHARGE
button on the front panel. Verify that a periodic tone sounds while the
defibrillator is charging. At the completion of
the charge cycle, the tone is continuous. The
charge time should be less than 7 seconds.
Verify that the
SHOCK
button (
) on the
front panel illuminates when the defibrillator
is charged.
11
Press and hold
on the front panel until
the unit discharges.
Verify that the defibrillator discharges and
that a
DEFIB SHORT TEST PASSED
message displays.
If the message
DEFIB SHORT TEST FAILED
displays, contact the appropriate technical
personnel or the ZOLL Technical Service
Department.
12 After the defibrillator display panel closes,
press the
PACER
button on the front panel to
display the Pacer Settings dialog box.
Verify that the Pacer Settings dialog displays
with the pacer status bar behind it.
13 Perform the following steps:
•
Set the Pacer mode to Fixed.
•
Set the Pacer Rate to 60 ppm.
•
Set the Output to 100 mA.
•
Select the Start Pacer button.
Verify that the green pace indicator on the
pacer status bar flashes once each second.
Verify that pace markers are printed every 25
mm on the chart recorder printout.
Note
: If a printout does not automatically
begin after starting the pacer, select the Print
quick access key to start a continuous
printout. Press the Print quick access key
again to stop the printout.
14 Press the
PACER
button on the front panel
button again to close the Pacer Settings
dialog box.
Verify that the pacer status bar indicates
PACING: SHORT DETECTED
.
Verify that the green pace indicator continues
to flash once each second.
Verify that no pacer-related error messages
display.
15 Disconnect the test connector from the MFC.
Verify that the pacer status bar displays:
PACING: CHECK THERAPY ELECTRODES
16 Press the
PACER
button on the front panel to
display the Pacer Settings dialog. Select
Turn
Pacer Off
to turn off the pacer.
Verify that the Pacer Settings dialog is closed
and that the pacer status bar is removed from
the display.
17 Reconnect the external AC Adapter to the X
Series rear panel.
Verify that the green Auxiliary Power indicator
light on the X Series front panel is
illuminated.
Function
Response
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...