![background image](http://html1.mh-extra.com/html/zoll/x-series/x-series_operators-manual_4055648079.webp)
Changing the Trends Status Window Display
9650-001355-01 Rev. M
X Series Operator’s Guide
4-5
Continuous Waveform Recording
Continuous waveform recording allows you to record continuous waveforms to a full
disclosure case log for the patient being monitored. This feature needs to be enabled by a
supervisor. The supervisor can access this feature by pressing the
Setup
quick access key (
),
and then selecting Supervisor>Log>Waveform Recording. In this menu, the following settings
can be customized:
•
Record Displayed Waveforms
– Record only the top waveform that is displayed or record
all waveforms that are displayed.
•
Record Additional Waveforms
– Always record the CO
2
waveform or always record the
IBP and the CO
2
waveforms.
There is an additional waveform that can be enabled in the Waveform Recording menu. This
option,
Record Pads Imp. Waveform
, measures the patient’s impedance between
defibrillation electrodes.
Note:
When Continuous waveform recording is disabled, no waveforms (except snapshots)
are recorded.
The full disclosure case can, at a minimum, concurrently store the following information:
•
32 monitor snapshots
•
500 non-ECG events
•
24 hours of continuous ECG (4 waveforms), Capnography, IBP (3 channels), and Pads
Impedance
The actual information that is stored depends on usage. Also, the specific combination of stored
continuous waveform data depends on how the waveform recording settings are configured in
the Supervisor menu.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...