Chapter 21 Patient Data
21-2
www.zoll.com
9650-001355-01 Rev. M
Storing Data
The X Series unit continuously stores information in a full disclosure case log for the patient
being monitored. The X Series can maintain up to 150 full disclosure cases that contain
treatment events, trends, ECG and other continuous waveforms, monitoring and event
snapshots, and 12-lead snapshots and analysis. The X Series unit can, at a minimum,
concurrently store the following information:
•
32 monitor snapshots
•
500 non-ECG events
•
24 hours of continuous ECG (4 waveforms), Capnography, IBP (3 channels), and Pads
Impedance
The actual information that is stored depends on usage. Also, the specific combination of stored
continuous waveform data depends on how the waveform recording settings are configured in
the Supervisor menu.
Note:
The X Series unit retains stored cases even if you turn off or remove the auxiliary
power adapter.
Log Capacity Indicator
The Log quick access key has an indicator bar which shows an approximation of the current
level of data storage in the log.
When data storage reaches its capacity, the unit performs automatic log management by
deleting the oldest full disclosure case. If the log is not cleared or transferred, the X Series unit
continues to delete cases as needed to acquire storage capacity. For more information on
clearing or transferring the log, see page 21-4.
Capturing a Data Snapshot
Press the Snapshot button (
) on the front panel to capture a 24-second period of numeric and
waveform patient data. The unit captures 12 seconds proceeding and 12 seconds following the
button press.
The X Series unit can store a minimum of 32 snapshots, including:
•
Monitor snapshots
•
Defibrillator snapshots
•
Treatment snapshots
•
Pacer snapshots
•
Alarm snapshots
•
Presenting Rhythm snapshots
Note:
Presenting Rhythm snapshots are only taken at the start of a new patient. The snapshot
will not be retaken if the unit is shut off for less than two minutes.
LOG
LOG
Indicator bar
The log is full.
The log is empty.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...