Blood pressure measurement may be inaccurate if taken while accelerating or decelerating
in a moving vehicle.
•
If an NIBP measurement result is questionable or “motion” indication is displayed, repeat
the measurement. If the repeated measurement result is still questionable, use another blood
pressure measurement method.
•
Do not use the NIBP on cardiopulmonary bypass patients.
IBP
•
To ensure compatibility and electrical safety, accessory pressure sensors should comply with
ANSI/AAMI BP-22 and IEC 60601-2-34 for IBP or ANSI/AAMI NS28 for ICP.
•
Follow instructions supplied with any accessory pressure sensor regarding calibration and
removal of trapped air.
•
Avoid touching metal parts of any transducer while it is in contact with the patient.
•
Do not reuse any components that are labeled for single use only.
•
Transducers should be rated to withstand an accidental drop of at least a meter onto a hard
surface.
•
Transducers that are subject to immersion in liquids should be rated as watertight.
CO
2
•
During MRI scanning, the monitor must be placed outside the MRI suite. When the monitor
is used outside the MRI suite, EtCO
2
monitoring can be implemented using a long
FilterLine
®
which permits placement of the monitor outside the MRI suite.
•
When using the monitor with anesthetics, nitrous oxide or high concentrations of oxygen,
connect the gas outlet to a scavenger system.
•
Use only Oridion Microstream CO
2
sampling lines.
•
Microstream CO
2
sampling lines are labeled for single patient use only. Do not reuse
sampling lines.
•
If using the CO
2
Monitor for extended critical care, replace the airway adapter every 24
hours or when it becomes occluded.
•
CO
2
readings and respiratory rate can be affected by sensor application errors, certain
ambient environmental conditions, and certain patient conditions.
Respiration
•
Do not operate the X Series with any other monitor with respiration measurements on the
same patient. The two devices could affect the respiration accuracy.
•
The device should not be used as an apnea monitor.
Ferromagnetic Equipment
Biomedical equipment and accessories, such as ECG electrodes, cables, and oximeter probes
contain ferromagnetic materials. Ferromagnetic equipment must not be used in the presence of
high magnetic fields created by magnetic resonance imaging (MRI) equipment.
The large magnetic fields generated by an MRI device can attract ferromagnetic equipment
with an extremely violent force, which could cause serious personal injury or death to persons
between the equipment and the MRI device.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...