9.1.
ADP/General Communication Conditions
I cannot communicate with the RiverSurveyor system or ADP.
This could be either a hardware or software issue that affects the communications interface.
Note: Cables and connectors are one of the highest causes of problems with field equipment; you
should make every attempt to take good care of all cables and connectors.
•
Check the cables between the computer and the ADP, or between the computer and the ra-
dio modem used with RiverCat systems. RiverCat users may want to connect their ADP di-
rectly to the computer; this could help determine if there are modem-related problems.
•
Check the power cable to the ADP, or check the batteries inside the RiverCat to make sure
they are within specifications.
•
Make sure you have connected the serial cable to the correct serial port on the computer. In
most cases, the Communications port should be set to COM1 and the baud rate set to 9600.
•
Check for any memory-resident programs running on your computer that may be accessing
the serial port. These programs may interfere with other programs trying to access the com-
puter serial ports. The most common type of memory-resident programs that interfere with
serial communications are PDA or Pocket PC synchronization programs. Close these pro-
grams down if they are causing the problem.
•
Use the
SonUtils
program to verify what COM ports are available on your computer.
•
The
Status
field in the
Systems
dialog box (§
) for the ADP may display a message that
says Invalid – Click here for information. Clicking on this link will open a troubleshooting
dialog box, which will describe possible causes and solutions.
The ADP indicator light is constantly red or keeps turning red.
This could be either a hardware or software issue that affects the communications interface.
•
See previous answer.
Profiles are being lost or there is intermittent communication.
This could be either a hardware or software issue that affects the communications interface.
•
See previous answer.
RiverSurveyor reported an error when trying to update my settings or start data collection.
There is a problem communicating with the ADP.
The system may have responded with the message “Failed to locate prompt”.
•
Check the cables between the computer and the ADP, or between the computer and the ra-
dio modem used with RiverCat systems. RiverCat users may want to connect their ADP di-
rectly to the computer; this could help determine if there are modem-related problems.
•
) settings. Try restarting the RiverSurveyor
program, or turning off the internal recorder, or adjusting your profiling settings. If the
problem persists, contact technical support. The problem may reside in a firmware setting.
•
Use the
SonUtils
program to verify what COM ports are available on your computer.
The RiverSurveyor software will not start collecting data.
There is a problem communicating with the ADP.
The system may have responded with the message “Failed to locate prompt”.
Summary of Contents for SonTek RiverSurveyor
Page 2: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 ii ...
Page 4: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 iv ...
Page 14: ......
Page 24: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 12 ...
Page 26: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 14 ...
Page 74: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 62 ...
Page 82: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 70 ...
Page 148: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 136 ...
Page 162: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 150 ...
Page 166: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 154 ...
Page 168: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 156 ...
Page 182: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 170 ...
Page 194: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 182 ...