6-8
IM 704510-01E
6.4
Setting Inhibit
For a functional description, see 2.5.
Procedure
1.
Press the SAMPLE key to display the Sample menu.
Inhibit does not appear when the sampling mode is set to inter-symbol interference
analysis and dual measurement function is used.
Selecting the Polarity
2.
Press the Inhibit soft key to select the polarity.
Selecting the Inhibit Input Level
3.
Press the INPUT key to display the Input menu.
4.
Press the Setting soft key to display the Input Setting menu.
5.
Press the Inhibit soft key.
6.
Press the Level soft key to display the Level selection menu.
7.
Press the soft key corresponding to the desired level.
Explanation
Selecting the Polarity
Select from the following. When set to OFF, the inhibit input is invalid.
•
:
Inhibits measurements while a positive signal is being applied to the inhibit
input terminal.
•
:
Inhibits measurements while a negative signal is being applied to the inhibit
input terminal.
Note
The time in which the inhibit input is valid is within the maximum sampling time from arming
activation.
Input Method of the Inhibit Signal
Apply a signal to the terminal marked INHIBIT on the front panel.
• Input impedance:
1 M
Ω
(typical value*)
• Input coupling:
DC
• Inhibit level:
Using the above procedure, select 0 V, TTL (1.4 V), or
TTL/10 (0.14 V).
• Maximum input voltage:
40 V (DC+ACpeak)
• Minimum input pulse width: 30 ns
• Setup time:
30 ns (must precede the input signal by at least 30 ns for
the inhibit to be valid).
*
Typical value represents a typical or average value. It is not strictly warranted.