6-12
IM 704510-01E
6.6
Setting the Input Coupling, Input Impedance,
and Trigger
For a functional description, see 2.5.
Procedure
1.
Press the INPUT key to display the Input menu.
Selecting the Source Channel
2.
Press the Setting soft key to display the Input Setting menu.
3.
Press the ChA or ChB soft key.
Selecting the Input Impedance
4.
Press the Impedance soft key to select the input impedance.
Selecting the Input Coupling
5.
Press the Coupling soft key to select the input coupling.
Selecting the Trigger Mode
6.
Press the Trigger soft key to display the Trigger menu.
7.
Press the soft key corresponding to the desired trigger mode.
Setting the Trigger Level
8.
Press the Level soft key and then set the trigger level using the rotary knob or
numeric keys.
Explanation
Selecting the Source Channel
Select the channel, CH A or CH B, for setting the input conditions.
Selecting the Input Impedance
Select 50
Ω
or 1 M
Ω
.
Note
The input impedance of the external arming input (EXT ARM), external gate input (EXT
GATE), and inhibit input (INHIBIT) is fixed to 1 M
Ω
.
Select the Input Coupling
Select AC or DC.
Note
The frequency characteristics vary depending on the input impedance and input coupling
settings. For a description of the frequency characteristics, see page 15-1.