15-4
IM 704510-01E
Item
Specifications
Pulse A & A-to-B time
• Measurement range
interval measurement and
• Time stamp mode or inter-symbol interference analysis mode
pulse A
→
A-to-B time
Pulse width measurement: 6 ns to 20 ms, A-to-B time interval measurement: 0 ns to 20 ms
interval measurement
• Hardware histogram mode
Pulse width measurement: 6 ns to 3.2
µ
s, A-to-B time interval measurement: 0 ns to 3.2
µ
s
• Inter-symbol interference analysis mode
Pulse width measurement: 10 ns to 3.2
µ
s, A-to-B time interval measurement: 0 ns to 3.2
µ
s
• Measurement resolution
• Pulse width measurement
• When the polarity is set to
or
:
±
100 ps rms
*1
rising edge trigger error
*2
±
falling edge trigger error
*2
• When the polarity is
:
±
100 ps rms
*1
rising edge trigger error
*2
±
falling edge trigger error
*2
±
trigger level timing error
*3
• A-to-B time interval measurement
• When the slope is set to A
↑
B
↑
, A
↓
B
↑
, A
↑
B
↓
, or A
↓
B
↓
:
±
100 ps rms
*1
±
A input trigger error
*2
±
B input trigger error
*2
• When the slope is set to
A B
or
A B
:
±
100 ps rms
*1
±
A input trigger error
*2
±
B input trigger error
*2
±
trigger level timing error
*3
• Accuracy
*4
• Pulse width measurement
• When the polarity is set to
or
:
Measurement resolution
±
(frequency stability of the time base
×
measured value)
±
trigger
level timing error
*3
±
1-ns systematic error
• When the polarity is
:
Measurement resolution
±
(frequency stability of the time base
×
measured value)
±
1-ns
systematic error
• A-to-B time interval measurement
• When the slope is set to A
↑
B
↑
, A
↓
B
↑
, A
↑
B
↓
, or A
↓
B
↓
:
Measurement resolution
±
(frequency stability of the time base
×
measured value)
±
trigger
level timing error
*3
±
1-ns systematic error
• When the slope is set to
A B
or
A B
:
Measurement resolution
±
(frequency stability of the time base
×
measured value)
±
1-ns
systematic error
• Polarity/Slope
• When in time stamp mode or hardware histogram mode:
Select from A
&
A B
and AA
&
A B
• When in inter-symbol interference analysis mode:
Select from A
&A
↑
B
↑
, A
&A
↓
B
↑
, A
&
A B
, A
&A
↑
B
↓
, A
&A
↓
B
↓
, and
A
&
A B
• Continuous A-to-B time interval measurement condition: The time to the next A signal edge after
the A-to-B time interval measurement is greater than equal to 13 ns and the time from the
previous A signal edge is greater than or equal to 20 ns
Pulse width A & pulse
• Measurement range
width B measurement
• When in time stamp mode: 6 ns to 20 ms
and pulse width A
→
pulse
• When in hardware histogram mode: 6 ns to 3.2 ms
width B measurement
• When in inter-symbol interference analysis mode: 10 ns to 3.2
µ
s (20 ns to 3.2
µ
s for pulse width B)
• Measurement resolution
• When the polarity is set to
or
:
±
100 ps rms
*1
rising edge trigger error
*2
±
falling edge trigger error
*2
• When the polarity is
:
±
100 ps rms
*1
rising edge trigger error
*2
±
falling edge trigger error
*2
±
trigger level timing error
*3
• Accuracy
*4
• When the polarity is set to
or
:
Measurement resolution
±
(frequency stability of the time base
×
measured value)
±
trigger
level timing error
*3
±
1-ns systematic error
• When the polarity is
:
Measurement resolution
±
(frequency stability of the time base
×
measured value)
±
1-ns
systematic error
• Polarity: Select from
,
, and
(
only when in inter-symbol interference analysis mode)
*1 100 ps rms or the display resolution whichever is greater when in hardware histogram mode, inter-symbol interference
analysis mode, or time stamp mode using multi window display.
*2 The trigger error, A input trigger error, B input trigger error, rising edge trigger error, and falling edge trigger error defined by
the following equation.
X
2
+ W
2
+ E
2
n
SR
X: Input amplifier noise, W: Cross talk noise (0.01
×
the signal amplitude of the other channel [Vrms])
En: Noise in the signal being measured [Vrms], SR: Slew rate of the input signal [V/s]
*3 The trigger level timing error is defined by the following equation.
8mV
8mV
Slew rate of the start signal
Slew rate of the stop signal
Trigger level setting accuracy
Slew rate of the start signal
Slew rate of the stop signal
Trigger level setting accuracy
±
±
±
–
*4 Measured value under standard operating conditions as described in General Specifications after the warm-up time has elapsed.
15.2 Measurement Functions (Measurement Items)