2-27
IM 704510-01E
Explanation of Functions
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
Printer Output of the Screen Image (see section 11.12 for the operating procedure)
The image displayed on the screen can be printed on the built-in printer.
Initializing the Settings (see section 13.1 for the operating procedure)
The setup data can be reset to their factory default settings.
Calibration (see section 13.2 for the operating procedure)
Using the internal calibration signal, the offset voltage of the input amplifier and the
conversion factors of the time and voltage converters can be calibrated. Calibration is
performed when the ambient environment (temperature and humidity) changes
drastically.
Changing the Beep and Click Sounds (see section 13.7 for the operating procedure)
A beep sound is heard when an error occurs. The beep sound can be turned OFF. You
can also select to turn ON or OFF the click sound that is heard when the rotary knob is
turned.
Error Logging (see section 14.3 for the operating procedure)
When a data error such as a data overflow or a communication error occurs, a
(warning mark) is displayed at the upper right corner of the screen. In such case, the
message is logged. All the messages can be viewed collectively as an error log later on.
Self Test (see section 14.4 for the operating procedure)
If you are in doubt as to whether the instrument has malfunctioned, you can run a self-
test before contacting a YOKOGAWA dealer. You can run checks on the circuit board,
the keys, the display, and other functions.
2.8 Other Functions