8-11
IM 704510-01E
Detailed Display Setup
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
Explanation
Auto Scaling
The X-axis and Y-axis is automatically set to values that are appropriate for the
measured data. The following limitations are present depending on the combination of
the sampling mode and window mode.
TimeStamp
HardHist
ISI
Single window
Yes
Yes
–
Multi window
No
Yes for Y-axis
No (Yes for A-to-B interval measurement only)
Auto window
No
Yes for Y-axis
No
TimeStamp: time stamp mode, HardHist: hardware histogram mode,
ISI: inter-symbol interference analysis mode
Manual Scaling
X-axis and Y-axis scaling can be specified. The items that are specified vary depending
on the sampling mode and display format.
• Histogram display when in hardware histogram mode or time stamp mode
(multi window or auto window)
Selectable range of the X-axis center: –50 ns to 3.200000000
µ
s (25 ps steps)
X-axis span selection:
Select from 1.5, 3, 7.5, 15, 30, 60, 150, 300, 600 ns, 1.5,
3, and 6
µ
s
Y-axis maximum selection: 10, 20, 40, 100, 200, 400, 1000, 2000, 4000, 10000,
20000, 40000, 100000, 200000, 400000, 1e6, 1e7, 1e8,
and 1e9 (Lin)
1e1, 1e2, 1e3, 1e4, 1e5, 1e6, 1e7, 1e8, and 1e9 (Log)
Y-axis scale type selection: Select from Lin (linear) or Log (logarithmic).
• Histogram display when using single window in time stamp mode
Selectable range of the X-axis center:
–50 ns to 20.000000000
µ
s (25 ps steps)
X-axis span selection:
Select from 1.5, 3, 7.5, 15, 30, 60, 150, 300, 600 ns, 1.5,
3, 7.5, 15, 30, 60, 150, 300, 600
µ
s, 1.5, 3, 7.5, 15, and
30 ms
Y-axis maximum selection: 10, 20, 40, 100, 200, 400, 1000, 2000, 4000, 10000,
20000, 40000, 100000, 200000, 400000, 1e6, 1e7, 1e8,
1e9 (Lin)
1e1, 1e2, 1e3, 1e4, 1e5, 1e6, 1e7, 1e8, 1e9 (Log)
Y-axis scale type selection: Select from Lin (linear) or Log (logarithmic).
• Time variation display in time stamp mode
Selectable range of the X-axis minimum: 0 to 320.0000000
µ
s (100 ns steps)
X-axis span selection:
Select from 6, 12, 30, 60, 120, 300, 600
µ
s, 1.2, 3, 6, 12,
30, 60, 120, 300, 600 ms, 1.2, 3, 6, 12, 30, 60, 120, 300,
600 s
Selectable range of Y-axis center:
–50 ns to 20.000000000
µ
s (25 ps steps)
Y-axis span selection:
Select from 500 ps, 1, 2.5, 5, 10, 20, 50, 100, 200, 500 ns,
1, 2., 5, 10, 20, 50, 100, 200, 500
µ
s, 1, 2, 5, 10, 20 ms
The following limitations are present in manual scaling.
TimeStamp
HardHist
ISI
Single window
Yes
Yes
–
Multi window (nT)
Yes
Yes
Yes for X-axis center
Multi window (SUM)
Yes for Y-axis
Yes for Y-axis
–
Multi window (ALL)
Yes for Y-axis
Yes for Y-axis
–
Auto window (nT)
Yes for Y-axis
Yes for Y-axis
–
Auto window (SUM)
Yes for Y-axis
Yes for Y-axis
–
Auto window (ALL)
Yes for Y-axis
Yes for Y-axis
–
TimeStamp: time stamp mode, HardHist: hardware histogram mode,
ISI: inter-symbol interference analysis mode
For dual measurement function, scaling can be specified on either Meas1 or Meas2 or both (Link).
8.4 Changing the Scale Value of the Histogram and Time Variation Displays