14-10
IM 704510-01E
Instrument Settings
• TA720:
None
• TA1100:
• Function:
Frequency A
• Gate time:
1 ms
• Number of samples: 1
• Sampling mode:
FREE
• Display:
NUMERIC
• Input Setting CH A: DC coupling, ATT=OFF, 50
Ω
, trigger level = 0 V
• REFERENCE:
EXTERNAL
• Rb atomic oscillator: None
Adjustment Procedure
• Adjust the TA720 after 30 minutes of warm-up.
• Turn the REFERENCE ADJUST on the rear panel of the TA720, and adjust it so that
the read-out value of the frequency counter is within the following range.
9.999995 MHz to 10.000005 MHz (10 MHz
±
5 Hz, within
±
0.5 ppm)
14.5 Adjusting the Time Base