Setting the Sampling Mode and Measurement Function
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
Continuous Measurement Conditions
When the measurement edge interval of the CH A input signal is 20 ns or more and the
measurement edge of CH B to the measurement edge of the CH A is 13 ns or more,
measurements are made continuously.
CH A
CH B
CH A
CH B
CH A
CH B
20 ns or less
Continuous
measurement
Non-continuous
measurement
20 ns or more
Measured
When set to A
↑
A
↑
B
↑
20 ns or more
13 ns or less
13 ns or more
13 ns or more
Measured
Measured
Measured
Measured
Measured
Measured
Measured
Measured
Measured
Measured
Measured
Measured
When A-to-B
time interval
measurement
is not continuous
Note
•
Changing the measurement function clears the data that has been acquired to the memory up
to that point.
•
Since period A & A-to-B time interval measurement is not possible when the sampling mode
is set to inter-symbol interference analysis mode (ISI), Period&TI does not appear as a
possible selection in the Function selection menu.
5.6 Setting the Period A & A-to-B Time Interval Measurement