9-14
IM 704510-01E
When the trigger is set to Between (Space-Space) and the data of ChA and ChB
are offset
Analyzes the Mark data of ChB that is in between Space (3T)-Space (3T) of ChA by
specifying the analysis start point using the symbol (Sync Pattern) search.
3T
Space
Mark
Mark
3T
3T
3T
4T
3T
6T
11T
11T
14T
7T
3T
6T
3T
14T
7T
3T
6T
Space
Space
Space
Mark
Mark
Space
Mark
Space
Mark
Mark
8T
3T
11T
6T
3T
6T
11T
3T
Sync Pattern
Setting
Mode: ISI
Function: PWA
→
PWB
> Slope: A B
Display > Window: Window setting suitable for the measured signal
> Item: List (to set the Sync Pattern)
> Sync PatternSet the start pattern of ISI analysis
> Item: Histogram (to display the ISI analysis results)
> Sync: ON
> ISI Setting > Mode: Between
> Polarity:
> Trigger:
> : 3T
> : 3T
:+
Sync Pattern
CH A measurement
CH B measurement
Trigger
Trigger
Analyzes the applicable Mark data.
Histogram display of
all Mark data
Histogram display of
extracted Mark data
Set “analyze the Mark data
between two Spaces(3T)”
Statistical values of
the extracted Mark data
Statistical values of
all Mark data
Note
In the pulse width A
→
pulse width B measurement function in inter-symbol interference
analysis mode, the constant T value of the window setting for MEAS1 (measured results of
CH A) applies also to Meas2.
9.3 Examples of Inter-symbol Interference Analysis