Chapter 6 Setting Measurement Conditions
6.1
Setting the Gate
For a functional description, see 2.5.
Procedure
1.
Press the SAMPLE key to display the Sample menu.
Selecting the Gate Mode
2.
Press the GateMode soft key to display the Gate Mode selection menu.
3.
Press the soft key corresponding to the desired gate mode.
• When event gate is selected
4.
Set the number of events (Event Size) using the rotary knob or numeric keys.
• When time gate is selected
4.
Use the rotary knob or numeric keys to enter the gate time.
• When external gate is selected
Selecting the Input Signal Polarity
4.
Press the Polarity soft key to select the input signal polarity.
Selecting the Gate Level
5.
Press the INPUT key to display the Input menu.
6.
Press the Setting soft key to display the Input Setting menu.
7.
Press the Arm/Gate soft key.
8.
Press the Level soft key to display the Level selection menu.
9.
Press the soft key corresponding to the desired level.