3.4
Turning ON/OFF the Power Switch
Items to be checked before turning ON the power
• Is the instrument properly installed? See section 3.2, “Installing the TA720”
• Is the power cord properly connected? See section 3.3, “Connecting to the Power”
Location of the Power Switch and ON/OFF Operation
The power switch is located at the lower left section of the front panel. The power switch
is a push button. Press once to turn it “ON” and press again to turn it “OFF.”
OFF
ON
Note
Allow at least 10 s when turning ON the power switch after turning it OFF. Otherwise, the
screen may not show anything. In such case, turn OFF the power switch again, wait at least
10 s, and then turn ON the power switch.
Power ON Operation
The following initial operation is carried out when the power switch is turned ON. Front panel key
operations are disabled during this period. When the initial operation is complete, the display
screen (screen that appears when the DISPLAY key is pressed) appears.
• ROM Check
• DRAM Check
• SRAM Check
• SRAM Cal Check: Calibration data check
Note
If the instrument does not operate as described above when the power switch is turned ON,
turn OFF the power switch and check the following points.
•
Check that the power cord is plugged in properly.
•
Is the correct voltage coming to the power outlet? See section 3.3.
If the TA720 still fails to power up when the power switch is turned ON after checking these
points, it is probably a malfunction. Please contact your nearest YOKOGAWA dealer as
listed on the back cover of this manual for repairs.
For Making Accurate Measurements
In the installation condition indicated in section 3.2, allow the instrument to warm up for
at least 30 minutes after the power switch is turned ON
Shutdown Operation
When the power switch is turned OFF, the setup data is stored, but measured results
and statistical calculation results are not. Save necessary data beforehand such as to a
floppy disk (section 11.5 and 11.6).
Note
The lithium battery that is used to store data has a limited life span. If the lithium battery is low,
the TA720 no longer operates properly such as the improper indication of date/time, or inability
to store/recall measured data. If such symptoms appear, execute a self test (see section 14.4).
If a warning message “Low Battery” is indicated as a result of the self test, the lithium battery
must be replaced quickly. The user cannot replace the battery. For battery replacement,
contact your nearest YOKOGAWA dealer as listed on the back cover of this manual.