13-6
IM 704510-01E
13.3 10-MHz Output
For a functional description, see 2.8.
Connecting the Cable
Connect a BNC cable to the 10MHz OUT terminal on the rear panel.
Explanation
When the power is ON, a signal of the following specifications are output from the
10MHz OUT terminal at all times.
• Output frequency:
10 MHz (typical value*)
• Output impedance:
50
Ω
(typical value*)
• Output coupling:
AC
• Output level: 1 Vp-p when the input impedance on the receiving side is 50
Ω
.
*
Typical value represents a typical or average value. It is not strictly warranted.
50
Ω
470 pF
10-MHz output circuit
CAUTION
Do not apply external voltage to the 10MHz OUT terminal. Doing so can cause
damage to the TA720.