9-13
IM 704510-01E
Statistical Calculation and Inter-Symbol Interference Analysis
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
Pulse Width A
→
Pulse Width B Measurement
When the trigger is set to Combination (Space/Mark)
Analyzes the Mark data of CH B immediately before the Space (3T)
→
Mark (7T)
combination pattern of CH A
3T
Space
Mark
Mark
3T
3T
3T
6T
11T
7T
14T
7T
Space
Space
Space
Mark
Mark
Mark
?T
?T
?T
?T
?T
?T
?T
?T
?T
Setting
Mode: ISI
Function: PWA
→
PWB
> Slope: A B
Display > Window: Window setting suitable for the measured signal
> ISI Setting > Mode: Combination
> Polarity:
> Trigger:
> : 3T
> : 7T
> Target: Prev.
> Sync: OFF
:+
CH A measurement
CH B measurement
Trigger
Analyzes the applicable Mark data.
Histogram display
(Displays all Mark data and
extracted Mark data overlapped)
Set “analyze the Mark data
immediately before the
Space(3T) and Mark(7T)
sequence”
Statistical values of
the extracted Mark data
Statistical values of
all Mark data
Trigger
Note
In the pulse width A
→
pulse width B measurement function in inter-symbol interference
analysis mode, the constant T value of the window setting for MEAS1 (measured results of
CH A) applies also to Meas2.
9.3 Examples of Inter-symbol Interference Analysis