Chapter 8 Detailed Display Setup
8.1
Setting the Single Window
For a functional description, see 2.6.
Procedure
1.
Press the DISPLAY key to display the Display menu.
2.
Press the Window soft key to display the Window Parameter Setting dialog box.
3.
Turn the rotary knob to select Mode, and then press the SELECT key to select Single.
Selecting the Setup Item (only for dual measurement function)
4.
Turn the rotary knob to select Item, and then press the SELECT key to select
Meas1Meas1 or Meas2.
Setting Constant T
• Setting Constant T Directly
5.
Turn the rotary knob to select T Value, and then press the SELECT key to show
the value entry area.
6.
Enter the desired constant using the numeric keys.
• Setting Constant T by Frequency
5.
Turn the rotary knob to select Frequency, and then press the SELECT key to
show the value entry area.
6.
Enter the desired frequency using the numeric keys.
Copying Settings
7.
Turn the rotary knob to select Copy to Meas2 or Copy to Meas1, and then press
the SELECT key.
If you selected Meas1 in step 4, Copy to Meas2 appears; if you selected Meas2,
Copy to Meas1 appears.
Explanation
Window Mode: Single
This mode is used when performing analysis using a single window.
Setting Constant T
Constant T can also be set using the frequency, which is an inverse of constant T
rounded to a resolution of 25 ps.
• Selectable range of constant T: 1 ns to 250 ns (25 ps steps)
• Selectable range when setting constant T using the frequency: 4 MHz to 1000 MHz