13.2 Calibration
For a functional description, see 2.8.
Procedure
1.
Press the UTILITY key to display the Utility menu.
2.
Press the Calibration soft key to display the Calibration menu.
3.
Press the Exec soft key.
The calibration results are listed.
Explanation
Calibrated Items
The following items are automatically calibrated using the internal calibration signal.
• Offset voltage of the input amplifier (DC calibration)
• Conversion coefficient of the time-to-voltage converter (T/V calibration)
• Delay line that sets the phase between channels during A-to-B time interval
measurement
(Phase calibration)
• Measurement function (Function calibration)
Executing Calibration
Perform calibration when the operating environment (temperature, humidity, etc.) of the
TA720 changes drastically.
If the power is turned OFF, the calibration data is reset to factory default values (see
page 13-3).
Calibration Results
If the calibration is successful, PASS is displayed; if it fails, FAIL is displayed. If FAIL is
displayed, perform a self test (see section 14.4) to diagnose the TA720.
Note
When executing calibration, leave the input terminals (CH A and CH B) open.