6-4
IM 704510-01E
6.2
Setting the Sampling Interval
For a functional description, see 2.5.
Procedure
1.
Press the SAMPLE key to display the Sample menu.
Interval appears only when the sampling mode is set to time stamp and the
measurement function is set to period, A-to-B time interval (TI), or pulse width.
2.
Press the Interval soft key.
3.
Set the sampling interval using the rotary knob or numeric keys.
Explanation
Setting the sampling interval
Set the sampling interval in the following range.
0
µ
s to 1 s (resolution: 1
µ
s)
When 0
µ
s is selected, sampling is performed at the following interval.
• Single measurement function: 12.5 ns
Note
•
The sampling interval is fixed to 0
µ
s when in hardware histogram mode and inter-symbol
interference analysis mode.
•
The time resolution of the time stamp does not change even when the sampling interval is
changed. It is always 0.1
µ
s.
•
When auto window is enabled, the constant T type (T Type) is set to Estimated T, and the
sampling interval is set to a value greater than or equal to 1
µ
s, an error message appears
and Estimated T is changed to Measured T.