6-7
IM 704510-01E
Setting Measurement Conditions
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
Selecting the Arming Delay Type
Select from the following. Selection can be made only when using external arming.
• OFF:
Arming delay selection and setting is not possible. Arming delay is not
executed.
• Time:
Measurement starts after the specified time of delay.
Selectable range of delay time: 1.0
µ
s to 1.0000000 s (100 ns steps)
• Event: Measurement starts after the specified event size.
Selectable range of event size: 1 to 1000000 (resolution: 1)
The arming delay can be set separately for each channel for the measurement functions
of period A & period B and pulse width A & pulse width B (pulse width A
→
pulse width
B).
Note
•
The event that is delayed is the event size of the input signal of the measurement channel.
However, for time interval measurement, the event size of CH A input signal is used.
•
When the polarity/slope is set to
,
A B
, A
A B
, A
A B
, A
, or
B
and the arming delay type is set to Event, one event is equal to a delay of one period.
•
If arming delay is set to a value other than OFF, RestMode of block sampling cannot be
specified.
6.3 Setting the Arming