3-6
IM 704510-01E
3.3
Connecting to the Power
Before Connecting the Power
Make sure that you observe the following points before connecting the power. Failure to
do so may cause electric shock or damage to the instrument.
WARNING
• Before connecting the power cord, ensure that the source voltage matches the
rated supply voltage of the instrument and that it is within the maximum rated
voltage of the provided power cord.
• Connect the power cord after checking that the power switch of the instrument is
turned OFF.
• To prevent the possibility of electric shock or fire, always use the power cord
supplied by YOKOGAWA.
• Make sure to perform protective grounding to prevent the possibility of electric
shock. Connect the power cord to a three-pin power outlet with a protective
earth terminal.
• Do not use an extension cord without protective earth ground. Otherwise, the
protection function will be compromised.
Connection Procedure
1.
Check that the power switch on the front panel is OFF.
2.
Connect the power cord plug to the power connector on the rear panel. (Use the
power cord that came with the package.)
3.
Connect the plug on the other end of the power cord to the outlet that meets the
conditions below. The AC outlet must be of a three-pin type with a protective earth
ground terminal.
Item
Specifications
Rated supply voltage*
100 to 120 VAC, 200 to 240 VAC
Permitted supply voltage range
90 to 132 VAC, 180 to 264 VAC
Rated supply voltage frequency
50/60 Hz
Permitted supply voltage frequency range
48 to 63Hz
Maximum power consumption (when using the printer)
250 VA
*
The TA720 can use a 100-V or a 200-V system for the power supply. The maximum rated
voltage of the power cord varies depending on its type. Check that the voltage supplied to
the TA720 is less than or equal to the maximum rated voltage of the provided power cord
(see page ii) before using it.
Power cord
(included in the package)
3-pin outlet