Appendix 14 High-Voltage 100 kS/s, 16-Bit Isolation Module (with RMS) (701260) Specifications
Item
Specifications
Response time
Rising (0 to 90% of 10 div):
100 ms (typical
10
)
(only when observing
Falling (100 to 10% of 10 div): 250 ms (typical
10
)
RMS)
Bandwidth limit
Select from OFF, 10 kHz, 1 kHz, and 100 Hz
Cutoff characteristics: –12 dB/OCT (typical
10
)
Probe attenuation setting
Voltage probe:
1:1, 10:1, 100:1, and 1000:1
Current probe:
10 A:1 V (for the 701933) and 100 A:1 V (for the 701930/701931)
Compatible probes/cables Connection cable (for high voltage 1:1): Recommended 1
701901 (isolated type BNC-safety alligator clip adapter
×
2: For measuring 850 V
(DC+ACpeak) or less),
701954 (alligator clip (dolphin type) red/black 2-piece set) is required separately
Voltage probe (10:1 safety probe): Recommended 2
700929 (10:1 safety probe) .20 to 45 pF: For measuring 1000 V (DC+ACpeak) or less
Current probe (power can be supplied from the SL1400)
701930 (150 A), 701931 (500 A), and 701933 (30 A)
1 Value measured under standard operating conditions (section 18.11).
Withstand voltage: 3700 Vrms for 1 minute
Allowable transient surge voltage:
±
5200 Vpeak
(between earth and input)
Combined with the 700929
2
4
7
3
700929
Combined with the 701954
H
L
Direct input (cable that does not comply with the safety standards)
8
9
BNC
H
L
701901
701954
5
H
L
6
10 Typical value represents a typical or average value. It is not strictly warranted.
WARNING
• When applying high voltage using this module, use the 1:1 safety cable
(combination of 701901 and 701954) or the isolated probe (700929).
• Note that the overvoltage tolerance of the low and high sides of this module may
differ depending on the combination of the connected probe, measurement lead,
and clip.
• Do not apply input voltage exceeding the maximum input voltage, withstand
voltage, or allowable surge voltage.
• To prevent electric shock, be sure to furnish protective earth grounding of the
SL1400.
• To prevent electric shock, be sure to fasten the module screws. Otherwise, the
electrical and mechanical protection functions will not be activated.
• Avoid continuous connection under an environment in which the surge voltage
may occur.