6-11
IM 701240-01E
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
App
Index
Triggering
6.6
Setting the External Trigger (SIMPLE)
<For a description of this function, refer to page 2-17.>
Procedure
1.
Press
MENU
.
2.
Press the
TRIGGER
soft key.
3.
Press the
Trigger Cond. Setup
soft key.
4.
Press the
Select
soft key to select Simple.
Setting the Trigger Source
5.
Press the
Source
soft key. The trigger source selection menu appears.
6.
Use
jog SELECT
to select Ext.
Setting the Trigger Slope
7.
Press the
Slope
soft key to select or .
Setting the Hold Off
8.
Set the hold off time according to the procedures given in section 6.4.
Explanation
The external signal that is input through the TRIG IN terminal on the left side panel of the
SL1400 can be used to generate triggers.
Note
For details on the TRIG IN terminal, see section 14.1.
Trigger Source
Select Ext.
Trigger Slope
Select how the trigger source is to cross the specified level for activating the trigger from
the following two choices.
:
Activated when the trigger source changes from below the trigger level to above
the trigger level (rising).
:
Activated when the trigger source changes from above the trigger level to below
the trigger level (falling).
Hold Off
See section 6.4.