Waveform Acquisition
7.2
Setting the Acquisition Mode
<For a description of this function, refer to page 2-28.>
Procedure
1.
Press
MENU
.
2.
Press the
ACQ
soft key.
Setting the Acquisition Mode
3.
Press the
Mode
soft key.
4.
Press the
Normal
,
Envelope
,
Average
, or
Box Average
soft key.
You may not be able to select some modes depending on the trigger mode setting.
Setting the Acquisition Count
5.
Press the
Count
soft key.
You cannot set the acquisition count when the trigger mode is Single, Single (N), or Log.
6.
Turn the
jog shuttle
to set the acquisition count. Pressing
RESET
resets the
count to Infinite.
If the acquisition mode is Average and you set Count to Infinite, proceed to step 7.
Setting the Attenuation (When the acquisition mode is Average and Count
is Infinite)
7.
Press the
Weight
soft key.
8.
Turn the
jog shuttle
to set the attenuation.