IM 701240-01E
5-4
5.3
Setting the Voltage Range
<For a description of this function, refer to page 2-6.>
Procedure
Selecting the Channel
1.
Press
CH
.
2.
Press one of the soft keys from
CH1
to
CH16
.
For the displayed menus when this operation is executed, see section 5.1.
Setting the Voltage Range
3.
Turn the
RANGE
knob to set the voltage range.
Explanation
Voltage Range (When Variable Is OFF)
The voltage range adjusts the displayed amplitude of the waveform. Set the voltage that
corresponds to the area between the top and bottom edges of the waveform display area
(10 divisions) for each channel. Set the voltage in 1-2-5 steps as in 1 V
→
2 V
→
5V.
When Variable is set to OFF, the Position and V Zoom settings are valid.
Note
•
The specified range is displayed at the upper left of the screen. If you change the range while
waveform acquisition is stopped, the new range is displayed in parentheses at the upper left
of the screen. The new value takes effect the next time acquisition is started.
•
Changing the range while acquisition is stopped has no affect on cursor measurement and
automated measurement of waveform parameters. The displayed values are for the range
setting that were used when the measurement was made.
Selectable Range
The selectable range varies depending on the input module as follows:
Input Module
Selectable Range
701250 (HS10M12)
50 mV to 200 V
701251 (HS1M16)
10 mV to 200 V
701255 (NONISO_10M12)
50 mV to 200 V
701267 (HV (with RMS))
200 mV to 2 kV
701261 (UNIVERSAL)
50 mV to 200 V
701262 (UNIVERSAL(AAF))
50 mV to 200 V
701265 (TEMP/HPV)
1 mV to 100 V
701275 (ACCL/VOLT)
50 mV to 100 V
The values are those when the probe attenuation is set to 1:1. Multiply the above values by 10,
100, and 100 for probe attenuation of 10:1, 100:1, and 1000:1, respectively.
Display Position of Range
If you turn the RANGE knob while waveform acquisition is stopped, the top row of the
range display shows the range for the displayed waveform, and the bottom row shows
the new range. The new range takes effect the next time acquisition is started.
Top row:
Range corresponding to the displayed waveform*
Bottom row: Changed range*
* If the range corresponding to the displayed waveform and the
changed range are the same, only the bottom row is displayed.