9-16
IM 701240-01E
9.6
Recording X-Y Waveforms
<For a description of this function, refer to page 2-40.>
Procedure
Before carrying out the procedure below, select X-Y Recorder mode according to the
procedures in section 9.1.
1.
Press
MENU
.
2.
Press the
RECORD COND
soft key.
Setting the Sample Rate
3.
Press the
Sample Rate
soft key.
4.
Turn the
jog shuttle
to select the sample rate.
Setting the Interpolation Method
5.
Press the
Interpolation
soft key. The interpolation selection menu appears.
6.
Press the soft key corresponding to the desired interpolation method.
Setting the Auto Print Function
7.
Press the
Auto Print
soft key to select ON or OFF. If you select ON, the X-Y
waveform recording on the built-in printer starts automatically when the
waveform acquisition stops.
Sample
rate
Record
time
Selecting the X-Y Waveform to Be Displayed (Recorded)
8.
Press the
Print Setup
soft key. The print setup dialog box opens.
9.
Use
jog SELECT
to select XY1 to XY4, turn the display ON/OFF of
each X-Y waveform, and set the X trace and Y trace channels.
Setting the Graticule
Setting the Grid
10.
Use
jog SELECT
to select the type.
Setting Dark/Light
11.
Use
jog SELECT
to select dark or light.
Setting the Grid Width of the X- and Y-Axes
12.
Use
jog SELECT
to select the width.