High-Speed 10 MS/s, 12-Bit Non-Isolation Module (701255) Specifications
Item
Specifications
Standard operating
Temperature: 23
±
5
°
C
conditions
Humidity:
20 to 80%RH
After a 30-minute warm-up and after calibration
Effective measurement
Twice the measurement range
range
Number of input channels
2
Input coupling
AC, DC, and GND
Maximum sample rate
10 MS/s
Input format
Non-isolated, unbalanced
Frequency characteristics
1
(–3 dB point when a sine wave with an amplitude equivalent to 60% of the range is applied)
DC to 3 MHz
Voltage-axis sensitivity
When using 10:1 probe attenuation: 500 mV to 2 kV (1-2-5 steps)
range
1:1 ProbeFactor:
50 mV to 200 V range (1-2-5 steps)
Maximum input voltage (at a frequency of 1 kHz or less)
Combined with the 701940 (10:1):
2
600 V (DC+ACpeak)
Direct input (cable that does not comply with the safety standards):
3
200 V (DC+ACpeak) (as a value that meets the safety standard)
250 V (DC+ACpeak) (maximum allowable voltage, as a value that does not damage the
instrument when applied)
Vertical (voltage) axis accuracy
DC accuracy
1
50 mV to 200 V range:
±
(0.5% of range)
Input connector
BNC connector (metallic type)
Input impedance
1 M
Ω
±
1%, approx. 35 pF
Lower –3 dB point when
10 Hz or less (1 Hz or less when using the 701940)
AC coupled
Common mode rejection
80 dB (50/60 Hz) or more (typical
4
)
ratio
Residual noise level
±
400
µ
V or
±
0.6% of the range, whichever is greater (typical
4
)
(Input section shorted)
A/D conversion resolution
12 bits (1500 LSB/range)
Temperature coefficient
Zero point: 50 mV to 200 V range:
±
(0.05% of range)/
°
C (typical
4
)
Gain:
±
(0.02% of range)/
°
C (typical
4
)
Bandwidth limit
Select from OFF, 500 kHz, 50 kHz, 5 kHz, and 500 Hz
Cutoff characteristics: –18 dB/OCT (typical
4
)
Probe attenuation setting
Voltage probe: 1:1, 10:1, 100:1, and 1000:1
Current probe: 10 A:1 V (for the 701933) and 100 A:1 V (for the 701930/701931)
Compatible probes/cables Voltage probe (10:1 passive probe): Recommended
701940, 17 to 46 pF: For measuring 600 V (DC+ACpeak) or less
Current probe (power can be supplied from the SL1400)
701930 (150 A), 701931 (500 A), and 701933 (30 A)
High voltage differential probe (connect the GND cable provided with the probe to the SL1400 case)
700924 (1000:1, 100:1/1400 Vpeak): For measuring 1400 Vpeak or less
Connection cable (for low voltage 1:1)
366926 (non-isolated type BNC-alligator clip
×
2: For measuring low voltage less than or equal
to 42 Vpeak)
1 Value measured under standard operating conditions (section 18.11).
H
L
2
Recommended:
Combined with the 10:1 passive probe
(701940)
H
L
701255
Input
terminal
Earth
3
Direct input
(cable that does not comply with the safety standards)
GND is connected to the case potential.
GND is connected to the case potential.
GND
GND
4 Typical value represents a typical or average value. It is not strictly warranted.
18.12 Module Specifications