Horizontal and Vertical Axes
5.14 Setting the Temperature Measurement
<For a description of this function, refer to page 2-12.>
Procedure
Selecting the Channel
1.
Press
CH
.
2.
Press one of the soft keys from
CH1
to
CH16
in which the 701261
(UNIVERSAL), 701262 (UNIVERSAL(AAF)), or 701265(TEMP/HPV)) is
installed.
For the displayed menus when this operation is executed, see section 5.1.
Selecting the Input Coupling of the Thermocouple
3.
Press the
Coupling
soft key to select TC.
Selecting the Thermocouple Type
4.
Press the
Type
soft key to select the thermocouple from K to Au7Fe.
Setting the Display Range
5.
Press the
Upper
soft key.
6.
Turn the
jog shuttle
to set the upper limit.
7.
Likewise, set Lower.
Selecting the Temperature Unit
8.
Press the
Next 1/2
soft key.
9.
Press the
Unit
soft key to select
°
C or K.
Setting the RJC
10.
Press the
RJC
soft key to select OFF or ON.
Setting the Burnout
11.
Press the
Burn Out
soft key to select OFF or ON.
AC is not available on the 701265.