Gloss Meter

Instruction 

Manual

R7700

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for R7700

Page 1: ...Gloss Meter Instruction Manual R7700 www com information itm com 1 800 561 8187 ...

Page 2: ...6 Data Recall 7 Clear Memory Data 7 Auto Power Off 7 Charging the Battery 7 Accessories and Replacement Parts 7 Applications 8 Product Care 8 Product Warranty 8 Product Disposal and Recycling 9 Product Support 9 Introduction Thank you for purchasing your REED R7700 Gloss Meter Please read the following instructions carefully before using your instrument By following the steps outlined in this manu...

Page 3: ... provided by an authorized service center Keep the lens protected from bright light When drastically changing the temperature environment of the unit allow the unit time to calibrate to the new environment Features Compact design allows for one handed operation Multi angle measurement 20 60 85 Backlit LCD display Auto calibration Internal memory stores last 254 data points Built in rechargeable ba...

Page 4: ...m Battery Internal Memory Yes up to 254 datapoints Product Certifications CE RoHS Operating Temperature 32 to 104 F 0 to 40 C Storage Temperature 4 to 140 F 20 to 60 C Operating Humidity Range 0 to 85 Dimensions 5 5 x 1 8 x 3 140 x 45 x 75mm Weight 10 76oz 305g Instrument Description 1 Display 2 Measurement Button 3 Delete Button 4 Data Recall Up Button 5 Power Button 6 Angle Button 7 Calibration ...

Page 5: ... now ready for use Note It is necessary to inspect the condition of the calibration tile and instrument optics before each use Any dust or debris on the optic should be removed from the lenses using dry clean air the lenses must not be touched If there are any permanent marks or scratches on the lens the unit is no longer suitable for measuring and should be returned for repair to an authorized di...

Page 6: ...oss 10 GU If measurement is less than 10 GU change test setup to 85 Taking Single Measurements 1 Place the measurement aperture against the material 2 Press the ANGLE button to select the right measuring geometry 3 Take measurements by pressing the MEAS button 4 The reading that displays within 2 seconds is the gloss value at the selected angle position 5 Each result is automatically logged into m...

Page 7: ...he MEAS or ANGLE buttons Clear Memory Data 1 When in measuring mode you can delete the most recently logged data by pressing the DEL button 2 To clear the data in memory press the DEL button for 3 seconds The stored reading will become 000 Note If there are no values to delete the display will show Err1 Auto Power Off To preserve battery life the meter is programmed to turn off after 30 minutes of...

Page 8: ...ternal parts only Product Warranty REED Instruments guarantees this instrument to be free of defects in material or workmanship for a period of one 1 year from date of shipment During the warranty period REED Instruments will repair or replace at no charge products or parts of a product that proves to be defective because of improper material or workmanship under normal use and maintenance REED In...

Page 9: ...t your authorized REED distributor or REED Instruments Customer Service by phone at 1 877 849 2127 or by email at info reedinstruments com Please visit www REEDINSTRUMENTS com for the most up to date manuals datasheets product guides and software Product specifications subject to change without notice All rights reserved Any unauthorized copying or reproduction of this manual is strictly prohibite...