Appendix A
SCPI Command Reference
CALCulate Subsystem
4460 GSM System Option and 4468 EDGE System Option
Version 12.20
659
:CALCulate:AFANalyser:ALL:LIMit:UPPer[:DATA]
Syntax
:CALCulate:AFANalyser:ALL:LIMit:UPPer[:DATA]
<real1>,<real2>,<real3>,<real4>, <real5>,<real6>
Parameters
realx
are six floating point real numbers.
real1
: The minimum value is
0.0
, the maximum is
30.0
. The minimum resolution
possible is
0.0001
, the default value is
5.0
.
real2
: The minimum value is
0.0
, the maximum is
30.0
. The minimum resolution
possible is
0.0001
, the default value is
5.0
.
real3
: The minimum value is
-40.0
, the maximum is
40.0
. The minimum resolu-
tion possible is
0.0001
, the default value is
5.0
.
real4
: The minimum value is
0.0
, the maximum is
20,000.0
. The minimum res-
olution possible is
1.0
, the default value is
5,000.0
.
real5
: The minimum value is
0.0
, the maximum is
100.0
. The minimum resolu-
tion possible is
0.1
, the default value is
5.0
.
real6
: The minimum value is
0.0
, the maximum is
100.0
. The minimum resolu-
tion possible is
0.1
, the default value is
20.0
.
Description
Sets the
upper
limits for the six main audio frequency measurements with one com-
mand. The order of the floating point real numbers defines their meaning:
real1
represents the upper limit of the peak-to-peak AC voltage measurement; the
physical dimension of the number stated is V(pp).
real2
represents the upper limit of the RMS-valued AC voltage measurement; the
physical dimension of the number stated is V(rms).
real3
represents the upper limit of the RMS-valued measurement of an AC voltage
ripple on a DC voltage; the physical dimension of the number stated is V(rms).
real4
represents the upper limit of the audio frequency measurement; the physical
dimension of the number stated is Hertz.
real5
represents the upper limit of the distortion measurement, carried out on the
third harmonic of a sine wave. The physical dimension of the number stated is per-
centage.
real6
represents the upper limit of the SINAD signal-to-noise ratio measurement;
the physical dimension of the number stated is dB.
Query
The query form of this command is not supported.
Example
:CALC:AFAN:ALL:LIM:UPP 3,3,1,4000,5,100
Sets the upper limits for the audio measurements: 3 V for the peak-to-peak AC volt-
age, 3 V for the rms-valued AC voltage, 1 V for the rms-valued ripple voltage, 4 kHz
for the audio frequency, 5% for THD, and 100 dB for SINAD.