Appendix A
SCPI Command Reference
CALCulate Subsystem
4460 GSM System Option and 4468 EDGE System Option
Version 12.20
715
:CALCulate:EGPRs:RFTX:FREQuency:LIMit:STATe
:CALCulate:EGPRs:RFTX:FREQuency:LIMit:UPPer[:DATA]
:CALCulate:EGPRs:RFTX:FREQuency:LIMit:LOWer[:DATA]
:CALCulate:EGPRs:RFTX:FREQuency:LIMit:GSM
Syntax
:CALCulate:EGPRs:RFTX:FREQuency:LIMit:STATe <PredefExp>
Parameters
PredefExp
is one of the following predefined expressions:
ON|OFF
. Default is
ON
.
Description
This command switches the check of the measurement result values of the
MEAS:EGPRs:RFTX:FREQuency
measurement against their user-definable
limits either
ON
or
OFF
.
Query
There is no query form of this command available.
Example
:CALC:EGPR:RFTX:FREQuency:LIM:STAT ON
Switches the limit check on frequency measurements on.
Syntax
:CALCulate:EGPRs:RFTX:FREQuency:LIMit:UPPer[:DATA]
<real1>
Parameters
real1
is a floating point real number. The minimum value for each is
0
, the maxi-
mum is
100000
. The resolution is
1
and the default value is
90
.
Description
Sets the upper limit for the frequency error measurement (in Hertz).
Query
There is only a command form available.
Example
:CALC:EGPR:RFTX:FREQ:LIM:UPP 45
Sets the upper limit of the frequency error measurement result to
45 Hz
.
Syntax
:CALCulate:EGPRs:RFTX:FREQuency:LIMit:LOWer[:DATA]
<real1>
Parameters
real1
is a floating point real number. The minimum value for each is
-100000
,
the maximum is
0
. The resolution is
1
and the default value is
-90
.
Description
Sets the lower limit for the frequency error measurement (in Hertz).
Query
There is only a command form available.
Example
:CALC:EGPR:RFTX:FREQ:LIM:LOW -60.8
Sets the lower limit of the frequency error measurement result to
-60.8 Hz
.
Syntax
:CALCulate:EGPRs:RFTX:FREQuency:LIMit:GSM <real1>
Parameters
real1
is a floating point real number. The minimum value for each is
0
, the maxi-
mum is
100000
. The resolution is
1
and the default value is
90
.